Maintenance of emergency shower and eye/ face wash With operation

Maintenance of emergency shower, Purpose: Maintenance of emergency shower, The purpose of this SOP is to describe the operation and maintenance of Emergency Shower and Eye/Face wash in Quality Control Laboratory. Maintenance of emergency shower, Scope: This procedure is applicable for Emergency Shower and Eye/ Face wash used in Quality Control Laboratory of general block […]

Maintenance of emergency shower, Purpose:

Maintenance of emergency shower, The purpose of this SOP is to describe the operation and maintenance of Emergency Shower and Eye/Face wash in Quality Control Laboratory.

Maintenance of emergency shower, Scope:

This procedure is applicable for Emergency Shower and Eye/ Face wash used in Quality Control Laboratory of general block of XX Pharmaceuticals Ltd.

Definitions / Abbreviation:

N/A

Responsibilities:

The roles and responsibility is as follows:

Executive, QC

[][]To ensure that this procedure is followed.
[][]To maintain the records properly as per SOP.

Manager, Quality Control

[][]To ensure that this procedure is kept up to date.
[][]To confirm that the SOP is technically sound and reflects the required working practices.
[][]To arrange training on the SOP to all concerned personnel and to ensure implementation of the SOP after training.

Head of Quality Assurance

[][]Approval of the SOP.
[][]To ensure the overall implementation of the SOP.

Annexure:

N/A

Operating Procedure:

Performance Checking:

[][]Check the unit performance once in a year.
[][]Turn the valve on to full open position (activation of the unit). The eyewash nozzles shall have a dust protecting cover, which shall be automatically removed upon activation of the unit.
[][]Ensure that the water flow of the unit at least 1.5 Liters/minute.
[][]Check that the unit shall deliver flushing fluid to both eyes simultaneously at a velocity low enough to be non-injurious to the user.
[][]Check that the flushing fluid temperature shall be normal. Temperature in the range of 20-30°C (about 80-95°F) is considered suitable.
[][]Check the Push or Pull valve shall not take more than 1 second to start the water flow

Eye/ Face washing:

[][]Position the eye or face on the shower.
[][]Pull “ROD” valve to open the water flow immediately.
[][]Rinse the effected eye with abundant water to remove any residual portion form the eye.
[][]After flushing, push “ROD” valve to stop the water flow.
[][]Keep the unit stand by position for the next use

Emergency Showering:

[][]Position the body under the unit.
[][]Expose the specific infected portion or whole body (if injured the whole body) exactly near “Shower Head”
[][]Pull “ROD” valve to open the water flow immediately.
[][]Open the valve to the full position.
[][]Flush with the sufficient water to the injured part or whole body until the chemical residue is removed.
[][]After flushing, push “ROD” valve to stop the water flow.
[][]Keep the unit stand by position for the next use.

Maintenance:

If Shower and Eyewash shows any mechanical, electrical or any others problem, inform to the supplier or Engineering Department for maintenance.