Biosafety Cabinet Operating Procedure

Biosafety Cabinet; Purpose To establish the ideal operation of Bio-safety Cabinet to confirm that the air are free from any microorganisms & any particle in order to favor microbiological test. Biosafety Cabinet; Scope This SOP applies for operation of Bio-safety Cabinet, Model: AC2-4E1 in Microbiology Section at XX Pharmaceuticals Limited. Definitions BSC    : Bio-safety Cabinet […]


Biosafety Cabinet; Purpose

To establish the ideal operation of Bio-safety Cabinet to confirm that the air are free from any microorganisms & any particle in order to favor microbiological test.

Biosafety Cabinet; Scope

This SOP applies for operation of Bio-safety Cabinet, Model: AC2-4E1 in Microbiology Section at XX Pharmaceuticals Limited.

Definitions
  • BSC    : Bio-safety Cabinet
  • DOP   : Dioctyl Phthalate
  • HEPA : High Efficiency Particulate Air.
  • IPA     : Iso Propyl Alcohol.
  • PAO   : Poly alpha olefin.
Responsibilities

The roles and responsibility is as follows

Executive, Microbiology

Operation, cleaning & disinfection and performance checking of Bio-safety Cabinet

Manager, Microbiology
  • Ensure Operation, cleaning & disinfection, performance checking and application of sound technical information.
  • Review that SOP is technically informative.
Head of Quality Assurance

Take initiative to Approval of SOP

Procedure:
Instructions
  • Do not wipe HEPA filter side of BSC.
  • Do not turn on UV light without full cover of slash.
  • Move always gently during operation of BSC to minimize air turbulence.
  • Care should be taken that water do not enter into HEPA filter through working station of down flow.
Operating Procedure
  • Press UV light button to turn on the UV light.
  • Wait for 30 minutes.
  • Press light button to turn on.
  • Pull up front slash at 50% open position. Open position (%) will be indicated on the display.
  • If show alarm, adjust position of slash.
  • Press FAN button to start up and wait for 3 minutes.
  • After 3 minutes, fan will be automatically started and
  • Use 70% ethanol/IPA to wipe all the surfaces of workstation, except in front of HEPA filter.
  • Use 70% ethanol to wipe all materials (Glassware’s, containers and other articles) before bringing them inside the Bio-safety Cabinet.
  • Start the work.
  • Use 70% ethanol to wipe again entire platform after finishing work.
  • Press light button to turn off.
  • Pull down to cover front slash at 0% open position. Open position (%) will be indicated on the display.
  • Press UV light button to turn on the UV light
  • Wait for 30 minutes to disinfect the whole cabinet.
  • After 30 minutes, press UV light button to turn off.
  • Disconnect the power plug of the cabinet.
Performance Check
  • Check the performance of HEPA filter by monitoring the Particle count and Microbial count once in a year and prepare the report in Annexure-I & Annexure-II
  • Check efficiency of HEPA filter by DOP/PAO test once in six months and prepare the report in Annexure-I & Annexure-II
Maintenance
  • Inform to the supplier or Engineering Department for maintenance, if Bio-safety Cabinet shows any mechanical, electrical or any others fault.
  • After maintenance, check HEPA filter efficiency & monitor Particle Count and microbial count.