Calibration of Ultrasonic Bath with Operation & Cleaning

Calibration of Ultrasonic Bath, Purpose : Calibration of Ultrasonic Bath, The purpose of this SOP is to describe the operation, calibration and cleaning of Ultrasonic Bath (Model: Clifton DU-14), used for degassing liquids, dissolving solids in solutions and cleaning laboratory apparatus. Calibration of Ultrasonic Bath, Scope : This procedure is applicable for Ultrasonic Bath (Model: […]


Calibration of Ultrasonic Bath, Purpose :

Calibration of Ultrasonic Bath, The purpose of this SOP is to describe the operation, calibration and cleaning of Ultrasonic Bath (Model: Clifton DU-14), used for degassing liquids, dissolving solids in solutions and cleaning laboratory apparatus.

Calibration of Ultrasonic Bath, Scope :

This procedure is applicable for Ultrasonic Bath (Model: Clifton DU-14), installed in the quality control laboratory of XX Pharmaceuticals Limited.

Definitions / Abbreviation:

[][]SOP: Standard Operating Procedure.
QC: Quality Control.
[][]Sonication: The act of applying sound energy (usually of the ultrasound type) to agitate particles in a sample.

Executive, QC

[][]To ensure that the instructions of this procedure are correctly followed.
[][]To maintain the records properly as per SOP.
[][]To ensure cleaning of ultrasonic bath maintaining safety rules.

Manager, Quality Control

[][]To ensure that this procedure is kept up to date.
[][]To confirm that the SOP is technically sound and reflects the required working practices.
[][]To arrange training on the SOP to all concerned personnel and to ensure implementation of the SOP after training.
[][]Schedule calibration of the instrument at the defined intervals.

Head of Quality Assurance

[][]To ensure the overall implementation of the SOP.
[][]Approval of the SOP.

Procedure:

Precaution(s):

[][]Never put hands in the bath during operations.
[][]Always maintain the liquid level above the minimum level mark.
[][]Disconnect the power supply before moving, emptying or cleaning of bath.
[][]Never operate the unit when dry.
[][]Allow the liquid in the bath to cool down before draining.
[][]Never block or restrict ventilation slots.
[][]Do not place item directly on the bottom of the bath, always make use of the stainless steel basket to suspend the item.
[][]Do not use cleaning fluids which contain concentrations of halogen ions, low pH values or high temperatures when operating the bath for cleaning purposes as these will all increase the risk of corrosion.

Operation:

[][]Check the calibration sticker to ensure that the instrument is within due date of calibration.
[][]Fill the bath with purified water above the ‘MIN’ level mark.
[][]Switch on the bath using the ‘ON/OFF’ switch, located at the rear of the bath.
[][]Place the stainless steel basket into position.
[][]Switch ‘ON’ the heater and press ‘■’ menu button to display “SP1”. Set the desired temperature using ▲ or ▼ key button. Maximum setting 69° C.
[][]After setting temperature, display will flash the temperature four times and then it will be automatically set.
[][]Press ‘■’ menu button twice to display “t”. Set the time using ▲ or ▼ key button.
[][]After setting time, display will flash the time four times and then it will be automatically set or press ‘■’ menu button to set manually.
[][]Place the glassware containing sample onto the stainless steel basket, to prevent touching the bottom of the tank.
[][]To start sonication, press and hold ‘Ο’ button until the Time LED is blinking.
[][]Instrument will stop after set time finished.
[][]Wait until the operation process is completed.
[][]To stop sonication, press and hold ‘Ο’ button until the Time LED is off.
[][]Use suitable tongs to remove the hot glassware from the bath.
[][]Allow the bath to cool.
[][]After completion of work, switch off the bath.
[][]Finally switch off the instrument.

Calibration:

[][]Calibrate the water bath once in a year.
[][]Fill purified water in the bath above the ‘MIN’ level mark.
[][]Switch ‘ON’ the mains.
[][]Switch ‘ON’ the mains of instrument.
[][]Set the desired temperature. Follow operation procedure 7.1.5 to 7.1.6.
[][]Allow 30 minutes to equilibrate.
[][]Check the temperature using a calibrated thermometer/digital thermometer and record the temperature in the calibration information sheet for Ultrasonic bath (as per Annexure-I).
[][]Set the desired time. Follow operation procedure 7.1.7 to 7.1.8
[][]Check the time using a calibrated stop watch and record the time in the calibration information sheet for Ultrasonic bath as per Annexure-I.

Cleaning procedure:

[][]Clean the instrument once in a week or in between when water becomes dirty.
[][]Switch ‘OFF’ the instrument and switch ‘OFF’ the mains.
[][]Remove the basket from the ultrasonic bath.
[][]Wash the basket properly with purified water.
[][]Drain the water from the ultrasonic bath.
[][]Wipe the inner surface of the bath with a non-abrasive cloth moistened in water.
[][]Use 10% nitric acid on a cloth (wear suitable gloves) to remove any scale or deposits.
[][]Fill purified water in the bath above the ‘MIN’ level mark.
[][]Put the basket in to the bath.

Annexure:

Annexure-I: Calibration Information Sheet for Ultrasonic bath.