Digital Weighing Balance operation, accuracy check at packaging area

Digital Weighing Balance operation, Purpose: Digital Weighing Balance operation, To lay down the procedure for operation, accuracy check and cleaning of the weighing balances at the Packaging Area. Digital Weighing Balance operation, Scope: This SOP is applicable for 820 gm weighing balance for weighing of materials at the Packaging Area of XX Pharmaceuticals Ltd. Definitions […]


Digital Weighing Balance operation, Purpose:

Digital Weighing Balance operation, To lay down the procedure for operation, accuracy check and cleaning of the weighing balances at the Packaging Area.

Digital Weighing Balance operation, Scope:

This SOP is applicable for 820 gm weighing balance for weighing of materials at the Packaging Area of XX Pharmaceuticals Ltd.

Definitions / Abbreviation:

N/A.

Responsibilities:

[][]The Roles and Responsibilities are as follows:\

Operator

[][]To follow the laid down procedure

Executive, Packaging

[][]To monitor and check the cleanliness of area

Manager, Production

[][]To implement this correctly

Head of Quality Assurance

[][]To approve the Document

Annexure:

N/A

Procedure:
Precaution:

[][]Do not pour liquids on the weighing platform.
[][]Do not use solvents to clean the platform.
[][]Ensure balance is clean before use, including its surrounding area.
[][]Check and ensure the due date of calibration before use.
[][]Check and ensure the level indicator of weighing balance is in the middle, if applicable.
[][]Use hand gloves to hold standard weights for function check and calibration.

Daily Performance Check :

[][]Plug in the power cord.
[][]Level the balance using leveling feet so that air bubble is centered within circle of level indicator.
[][]To turn the balance ON, short pressing the → O/T ← key.
[][]Wait a while so that the stability indicator [*] is displayed.
[][]Zero the balance by pressing the → O/T ← key.
[][]Press the 100 mg mass at the centre of the platform and record the value displayed in the daily balance check logbook.
[][]Press the 100 g mass at the centre of the platform and record the value displayed.
[][]Press the 200 g mass at the centre of the platform and record the value displayed.
[][]Print the results by pressing print ( ) button if necessary.
[][]Record the value of those weights in Log Book for Daily Accuracy Check (As per Form No.: SOP/ PRO/004/I).
[][]If any displayed value is out of acceptable limit, label the balance “DO Not Use” and inform the QA Executive for immediate calibration of the balance.
[][]Balance check must be performed every working day before starting operation.
[][]If the factory remains close for the day(s), put the comments in the remarks column against that day(s) at the day of factory resume.

Operation of 820 g Balance of Mettler Toledo :

Empty container for Weighing:

[][]Press the zero button.
[][]Place an empty container on the scale platform.
[][]Press the Tare button.
[][]Fill the container with material to be weighed.
[][]The terminal will display the net weight and the NET cursor will light up. Allow the reading to be stable. Write down the reading.
[][]To clear a tare weight (with the scale in the net weight mode and a tare weight previously entered),press the Clear key (C)

Filled container for Weighing:

[][]Press the zero button.
[][]Place the material to be weighed on the scale platform.
[][]The terminal will display the net weight and the NET cursor will light up. Allow the reading to be stable. Write down the reading.
[][]Follow the Daily Accuracy Check.

Acceptance Criteria:

[][]The observed weights should be within ± 0.1% of the certified standard weights.
[][]If the acceptance criteria are not satisfied, follow the procedure for Handling of out of calibration for instrument & equipment.

Calibration:

[][]Perform the calibration in every six months. Recalibration is required in the following cases :
[][]After any minor service has been done or replacement of parts.
[][]When the software or firmware has been upgraded or changed.
[][]If daily check fails.

Cleaning:

[][]Place the status label ‘TO BE CLEANED’ on the equipment.
[][]Switch off the balance and unplug from the socket.
[][]Remove all objects from the weighing platform of balance.
[][]Lift up and remove the weighing pan in such a way that there is no damage in weighing system.
[][]Clean the pan including underside using a piece of dry cloth. Use wet cloth and finally clean with disinfection solution
[][]Wipe the display, switch board and cables with dry cloth.
[][]Remove ‘TO BE CLEANED’ and affix ‘CLEANED’  label on the equipment after approval from Quality Compliance.
[][]Record the cleaning parameters on the ‘Equipment Log Book’ .

Annexure:

Annexure I Equipment Use Log Book


Leave a Comment

Your email address will not be published. Required fields are marked *

error: Content is protected by www.pharmaguideline.co.uk !!