Maintenance of Desiccator

Maintenance of Desiccator; Purpose : Maintenance of Desiccator; The purpose of this procedure is to describe the maintenance (handling and cleaning) of desiccator. Scope : This procedure is applicable for all desiccators in the quality control laboratory of XX Pharmaceuticals Ltd. Definitions / Abbreviation: Desiccant canister: A container with hygroscopic substance which induces or sustains […]


Maintenance of Desiccator; Purpose :

Maintenance of Desiccator; The purpose of this procedure is to describe the maintenance (handling and cleaning) of desiccator.

Scope :

This procedure is applicable for all desiccators in the quality control laboratory of XX Pharmaceuticals Ltd.

Definitions / Abbreviation:

Desiccant canister: A container with hygroscopic substance which induces or sustains a state of dryness in its area.

Responsibilities:

Officer/ Executive, QC

[][] To follow the instructions of this procedure correctly.

Sr. Executive, QC

[][] To ensure that SOP is followed correctly.

Manager, Quality Control

[][] To ensure that this procedure is kept up to date.
[][] To ensure appropriate personnel from the section are trained on this procedure.
[][] To confirm that SOP is technically sound and reflects the required working practices.

Manager, Quality Assurance

[][]Approval of the SOP.
[][]To ensure the overall implementation of the SOP.

Procedure:

Precaution(s):

[][]Do not touch hot samples, while transferring them from Oven/Muffle furnace to desiccator.

Operation

[][]Place about 200 g of silica (desiccant) and/or dessicant canister into the bottom of the desiccator.
[][]Set the sample platform (desiccator plate) to the desiccator.
[][]Apply a smear of grease in between lid & desiccator rim to avoid sticking of lid with the dessicator.
[][]Close the lid properly.
[][]Open the lid to retain the sample to be analyzed and close the lid again.
[][]Remove the sample after required period of time.

Cleaning

[][]Rub the desiccator, sample platform (desiccator plate) and lid with a dry, clean cloth.
[][]Laboratory Attendant will collect the silica (desiccant) to activate (in the oven at about 105°C) and then replace it to the desiccator once in a week or even more frequently when blue color of desiccant turns purple/pink.
[][]Fill up the logbook for desiccant change (as per Annexure-I).
[][]Officer/Executive, QC will check the logbook for proper condition of desiccant or the desiccant change.

Annexure:

Annexure-I: Logbook for desiccant change of desiccator.