Oven Operation, cleaning and calibration procedure

Oven Purpose To dry all properly cleaned glassware which is subject to use in chemical analysis & also to sterilize cleaned glassware which is subject to use in different microbiological analysis. Oven Scope This SOP applied for operation of Oven, which model is: UNE600 in QC Section, Microbiology Section & Product Development Section of at […]


Oven Purpose

To dry all properly cleaned glassware which is subject to use in chemical analysis & also to sterilize cleaned glassware which is subject to use in different microbiological analysis.

Oven Scope

This SOP applied for operation of Oven, which model is: UNE600 in QC Section, Microbiology Section & Product Development Section of at XX Pharmaceuticals Limited.

Definition/Abbreviation

N/A

Responsibilities

The roles and responsibility is as follows

Executive, Microbiology/ Executive, QC

Operation, cleaning & disinfection and performance checking of oven.

Assistant Manager, Microbiology
  • Ensure Operation, cleaning & disinfection and performance checking.
  • Ensure that oven is calibrated.
  • Checking that SOP is technically informative.
Head of Quality Assurance

Take initiative to Approval of SOP

Procedure
Instructions
  • Never enter open hands in the chamber of oven.
  • To keep or remove glassware, wear temperature resistance gloves.
  • Avoid to keep glassware into chamber with water or any other solution.
  • Do not dry plastic materials using oven.
  • Always keep door closed & avoid opening door for long period.
Operating Procedure
Sterilization of Glassware for Microbiological Analysis
  • To put on the main power switch, Press push/turn control key in front of the instrument. Oven will start in normal mode with the display of the chamber temperature, timer, alarm temperature.
  • Turn the push/turn control key by hold down the SET key & at the clockwise or anti clockwise for setting date, local time, operating temperature, alarm temperature. After setting, SET key will be released the display briefly flashes the set point.
  • To select the programme, Press & hold down SET key and rotate push/turn control key at clockwise and select “Ramp timer” mode.
  • Set the work days group using the push/turn control by holding down the SET key and Press push/turn control to select the ramp segment “t1” If the day is not required, it will be off position.
  • Set the time using push/turn control by hold down the set key &
  • Use the push/turn control, to Select ramp segment “t2”. Hold down SET key and using the push/turn control set the time.
  • Until the temperature display is flashing, Turn push/turn control clockwise. Set the required temperature setpoint using push/turn control
  • Select setpoint waiting time by hold down the SET key. Hold down SET key and set on using the push/turn control. After the SET key has been released the function for setpoint waiting time is stored.
  • Hold down SET key & set time using the push/turn control then Select hold time “t3”.
  • Hold down SET key & set time using push/turn control after selecting cooling time “t4”.
  • Select program repeats “loop”. Hold down the SET key and set 2 for repeats using the push/turn control.
  • To set the air changes, select the fan speed. Turn the push/turn control clockwise until the fan symbol is flashing. Set fan speed 50% suing the push/turn control holding down the SET key.
  • Turn the push/turn control clockwise until monitor temperature display is flashing. Set the alarm temperature using the push/turn control holding down the SET key & select alarm temperature.
  • Turn push/turn control at clockwise until the fan symbol flashing to move the air slider opens & closes the air valve to control the supply and discharge of air.
  • Turn the push/turn control clockwise until the stop symbol ▀  is flashing. Select start ► using the push/turn control holding down the SET key and On releasing the SET key the program starts to run
Glassware Drying
  • To put on main power switch. Press push/turn control key in front of the instrument. The oven will start in normal mode with display of chamber temperature, timer, alarm temperature.
  • Select operating mode “Normal operation” holding down SET key (approximately 3 seconds) the current operating is flashing.
  • Select temperature set point by hold down the SET key & use the push/turn control to select required temperature set point. After the SET key has been released the oven briefly flashes the temperature setpoint. Heating is indicated by the orange heater symbol.
  • Select fan speed. Turn push/turn control clockwise until fan symbol is flashing while the holding down the SET key. To set 50% fan speed use the push/turn control
  • Turn the push/turn control clockwise until the monitor temperature display is flashing. Select the alarm temperature. Use the push/turn control to set the alarm temperature by holding down the SET key.
  • The display then changes to the actual current temperature and starts to the setting temperature. The temperature will be automatically increased at setting temperature and display the setting temperature digitally.
  • Observe the display temperature until stable position.
  • Remove dried glassware from oven at the end of drying,
Cleaning of Oven
  • Clean the inner benches of Oven with a cleaned cloth.
  • Use 70% IPA or ethanol to disinfect the inner surfaces.
  • Clean the outer surfaces with dry cleaned cloth.
  • Disinfect the outer surfaces with 5% Dettol /Savlon solution.
Calibration of Oven
  • Insert standard thermocouples or thermometer into the chamber.
  • At least 6 position to be consider to place different location of chamber.
  • Press & hold down SET key until the normal mode light blinking.
  • To reach SETUP light, turn push/turn control key clock wise.
  • Select Calibration to rotate push/turn control key.
  • Use the following temperature for calibration:
  • 1 : 1500C
  • 2 : 2000C
  • 3 : 2500C
  • Select required calibration temperature in SETUP key & set the corresponding calibration correction to 0.00
  • Under the steady conditions, using a reference instrument, measure the deviation from the selected calibration temperature.
  • If the measured reference temperature is too low, the calibration correction setting is negative sign, set the calibration correction in SETUP key.
  • The correction value should not more than ±10C.
  • Adjust the temperature if require from adjusting value of calibration temperature (CAL), carry out the check measurement using the reference thermometer.
  • Carry out others two calibration temperatures in the same manner.
  • Calibrate incubator once in a year.
  • Record calibration in Oven Calibration Record, Annexure-II.

Maintenance

Inform to supplier or Engineering Department for maintenance, If Oven shows any mechanical, electrical or any others problem, after maintenance, recalibrate the oven.

Download all Annexure here

Annexure I Oven Calibration Record

Annexure II Oven Log Book

 

 