Vortex Mixer; Operation and Cleaning Procedure

Vortex Mixer; Purpose To establish the procedure for ideal operation and cleaning of Vortex Mixer, Model: Genius 3 KIA, Fisher Scientific, UK in microbiology test. Vortex Mixer; Scope This SOP applies for operation, cleaning of Vortex Mixer, Model: Genius 3 KIA, Fisher Scientific, UK in Microbiology Section at XX Pharmaceuticals Limited. Definitions/Abbreviations NA Responsibilities The […]


Vortex Mixer; Purpose

To establish the procedure for ideal operation and cleaning of Vortex Mixer, Model: Genius 3 KIA, Fisher Scientific, UK in microbiology test.

Vortex Mixer; Scope

This SOP applies for operation, cleaning of Vortex Mixer, Model: Genius 3 KIA, Fisher Scientific, UK in Microbiology Section at XX Pharmaceuticals Limited.

Definitions/Abbreviations

NA

Responsibilities

The roles and responsibility is as follows:

Executive/ Senior Executive, Microbiology

Operate & clean of Vortex Mixer.

Asst. Manager, Microbiology

  • Ensure Operation, cleaning & application of sound technical information.
  • Keep the up to date of SOP.
  • Review that SOP is technically informative.
Head of Quality Assurance

Take initiative to Approval of SOP

Procedure:
Instructions
  • Do not press hardly on platform.
  • To avoid movement during vortex, keep Vortex Mixer in a fixed place.
  • Do not vortex any materials for long time.
  • To avoid overflow during vortex, maintain the level of test solution into test tube/Vial/test flask.
 Operation Procedure
  • Switch ON the power button. The red indicator light will be illuminated.
  • Press SET switch to select the automatic vortex or manual vortex.
  • Set time to press TIMER button as per requirement.
  • Adjust vortex speed by turning SPEED button at clockwise for increasing or anti clockwise for decreasing.
  • Place test tube/vial/test flask on the platform.
  • Attach tube/vial with the machine for more than 1 minute vortex.
  • Press START button if set manual vortex.
  • Touch test tube/vial on platform to start the vortex automatically.
  • Remove test tube at the end of vortex.
  • Switch OFF power button.
  • Disconnect machine.
  • Press MAINS switch off.
Cleaning Procedure
  • Switch off Vortex Mixer and disconnect the power plug.
  • Clean outer surface with dry cloth.
  • Disinfect outer surfaces of Microscope with 5% Dettol/Savlon solution.
  • Clean daily once.

Maintenance

Inform to supplier or Engineering Department for repairing, If Colony counter shows any mechanical, electrical or any others problem.