Water Bath, Operation, Calibration and Cleaning

Water Bath; Purpose The tenacity of this SOP is to define the operation, calibration & cleaning of Water Bath Water Bath; Scope This procedure is applicable for Water Bath installed in Microbiology section at XX Pharmaceuticals Limited. Definitions/Abbreviations LED: Light Emitting Diode Responsibilities: The roles and responsibility are as follows Executive/Senior Executive, Microbiology To confirm […]


Water Bath; Purpose

The tenacity of this SOP is to define the operation, calibration & cleaning of Water Bath

Water Bath; Scope

This procedure is applicable for Water Bath installed in Microbiology section at XX Pharmaceuticals Limited.

Definitions/Abbreviations

LED: Light Emitting Diode

Responsibilities:

The roles and responsibility are as follows

Executive/Senior Executive, Microbiology
  • To confirm that the instructions of this procedure are properly followed.
  • To maintain the record correctly as per SOP.
  • To confirm cleaning of shaking water bath maintaining safety guidelines.
Assistant Manager, Microbiology
  • To confirm that this process is kept up to date.
  • To ensure that the SOP is technically sound & reflects required working practices.
  • To arrange training on the SOP to all concerned personnel & to ensure implementation of the SOP after training.
  • Schedule calibration of the instrument at the defined intervals.
Procedure
Instructions
  • Laboratory coat, hand gloves & safety glasses must be worn while handling instrument.
  • Never touch the liquid within the bath as it may be very hot.
  • Each time use the display or a thermometer to check the temperature.
  • Never block ventilation slots during operation.
  • Each time disconnect bath from electricity supply before cleaning.
  • Allow the liquid in bath to cool down to 40°C before draining.
  • Use only purified water, never operate water bath without water.
  • Fill the it prior to connection to power supply.
  • It should be covered with a lid or polypropylene spheres to achieve optimum performance.
  • Always use the stainless steel flat lid with concentric rings when the instrument is not in use to avoid contaminants landing in the bath liquid.
  • Don’t touch any electrical contacts or open any closure panel that may be risk of electrical shock when above 60˚C or below room temperature,
Operating Procedure
  • Check the calibration date of the instrument before each time use. If the calibration is out of date, calibrate it and then use.
  • Check water level, If require adjust the water bath to an appropriate level with purified water.
  • Switch on the power from the main.
  • Using the switch “I” ON, “O” OFF button located at the rear side of it, Switch on the Water bath When ON (I) the switch is illuminated & unit performs a self-test where all segments of the 3 digit LED display & indicators illuminated.
  • To display “SP1” to set temperature, press & hold down key for more than 1.5 seconds. “SP1” will show blinking.
  • To select required temperature, use up or down arrow keys.
  • Heater indicator will illuminate. Wait until the display flashes between “SP1” & set temperature and will revert to show actual liquid temperature.
  • When the bath temperature is either 4°C above or below set temperature, LED indicator will illuminate.
  • Open concentric ring & place beaker or suitable container with sample on sphere of it.
  • After completion of work, switch off bath by using “I” ON, “O” OFF switch, located at rear side of back of it.
Calibration Procedure
  • Calibrate Water bath once in a year.
  • Fill purified water in water bath to an suitable level.
  • Switch on the bath using “I” ON, “O” OFF switch, located at the rear side of back of the Water bath.
  • Set desired temperature against calibrated standard thermometer at 35⁰C, 45⁰C, 65⁰C.
  • Wait 30 minutes to equilibrate.
  • Verify temperature using a calibrated standard thermometer & record the temperature in the Calibration information sheet for Water bath (as per Annexure-I).
Cleaning Procedure
  • When water becomes dirty, clean instrument otherwise clean once in a week.
  • Switch ‘OFF’ the instrument.
  • Disconnect main power supply.
  • Open stainless steel flat lids with concentric rings of the water bath.
  • Use purified water to wash the stainless steel flat lids and concentric rings.
  • To reduce potential biological contamination, remove the water from water bath.
  • Wash it with soapy water & then purified water.
  • Mop with clean dry cotton cloth.
  • To remove any deposits, use 10% nitric acid on a cloth (wear suitable gloves).
  • Gently heat to 50°C for an hour after adding 1 liter of vinegar to water in the stainless steel tank, empty and brush the lime away for descaling then rinse thoroughly afterwards.
  • Fill with purified water up to desired level .
  • Close the stainless steel flat lids with concentric rings.

 Download Annexure Here

Annexure I Calibration Information Sheet for Water Bath


Leave a Comment